Fresenius Kabi USA, LLC: Ivenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004.
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Fresenius Kabi USA, LLC
- Product
- Ivenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004.
- Reason / hazard
- Downstream Occlusion alarms may occur during procedures using low flow infusion rates when a check valve* is added to the Ivenix Administration Set.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the sates of CA, CO, FL, GA, ID, IL, MD, MI, MN, MS, NE, NJ, NV, OK, OR, SC, TX, UT, VA, WA, WI.
- Recall initiated
- 2025-11-03
- Code info
- Model Number: LVP-0004. UDI-DI: 00811505030320.
- Product quantity
- 15,862 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- North Andover
- State
- MA
- Country
- United States
- Event id
- 97939