Medline Industries, LP: MEDLINE convenience kits labeled as: 1) BASIC NEURO PACK, REF CDS780202R; 2) ACDF CDS, REF CDS941061M; 3) NEURO PACK-LF, REF DYNJ0038
Read the official notice at FDA (device) →
Get alerts for Medline Industries
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medline Industries, LP
- Product
- MEDLINE convenience kits labeled as: 1) BASIC NEURO PACK, REF CDS780202R; 2) ACDF CDS, REF CDS941061M; 3) NEURO PACK-LF, REF DYNJ0038440C; 4) NEURO VP SHUNT/VAGAL NERVE-RF, REF DYNJ21925S; 5) NEURO-ORTHO EXTRAS, REF DYNJ44882M; 6) LAMINECTOMY, REF DYNJ903870F.
- Reason / hazard
- Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution.
- Recall initiated
- 2025-09-30
- Code info
- 1) REF CDS780202R, UDI/DI 10195327418618 (each) 40195327418619 (case), Lot Numbers: 23LMG063; 2) REF CDS941061M, UDI/DI 10193489445589 (each) 40193489445580 (case), Lot Numbers: 24ABG356, 24CBH117, 24DBJ347; 3) REF DYNJ0038440C, UDI/DI 10195327356699 (each) 40195327356690 (case), Lot Numbers: 23JMC956, 23KMF585, 24AMB974; 4) REF DYNJ21925S, UDI/DI 10195327105020 (each) 40195327105021 (case), Lot Numbers: 23EBS176, 23GBF899, 23HBZ085; 5) REF DYNJ44882M, UDI/DI 10195327382803 (each) 40195327382804 (case), Lot Numbers: 23FDA522, 23GDB903; 6) REF DYNJ903870F, UDI/DI 10195327239268 (each) 40195327239269 (case), Lot Numbers: 23IBD913, 23IBD914, 23JBL889, 24ABE596, 24BBL042, 24CBH196, 24DBU825.
- Product quantity
- 1315 kits
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Northfield
- State
- IL
- Country
- United States
- Event id
- 97846
Free daily digest + "My products" watchlist