⚠ RecallRadar📷 ScanGuidesRecentOfficial U.S. recall notices, restated plainly

restor3d Inc.: Generic Name: restor3d MTP Implant Generic Name: MTP Implant Model/Catalog Number: 1103-161620 Software Version: None Product Descriptio

Medical DevicesSource: FDA (device) · Date: 2026-10-28

Recall status
Ongoing (still active)
Severity
Class II: could cause temporary or reversible health problems What recall classes mean
Where sold
U.S. Nationwide distribution in the states of FL, IL, NC, PA, SC, and TX.

Common questions

Is restor3d MTP Implant recalled?
Yes. FDA (device) recalled restor3d MTP Implant on 2026-10-28. Recall status: Ongoing (still active).
What is the hazard?
Implants were found in the field with orange sterilzation indicators which means implants were not exposed to sterilization.
Where was it sold?
U.S. Nationwide distribution in the states of FL, IL, NC, PA, SC, and TX.

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Is restor3d recalled?

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

More recalls

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
restor3d Inc.
Product
Generic Name: restor3d MTP Implant Generic Name: MTP Implant Model/Catalog Number: 1103-161620 Software Version: None Product Description: The restor3d Metatarsophalangeal (MTP) Implant is a sterile, single-use implant grade metal device, available in varied articulating surface coverages. The restor3d MTP implant is designed to replace the damaged articulating surface of the first metatarsal distal head. The device is intended for the treatment of patients with degenerative and post-traumatic arthritis in the first metatarsal joint in the presence of good bone stock and integrity of the first metatarsal head. This implant can also be used in the following clinical conditions: hallux valgus or hallux limitus, hallux rigidus, and an unstable or painful metatarsal/phalangeal (MTP) joint. The implant is a single use product intended to be press fit, with optional use of bone cement. The restor3d MTP Implant is comprised of a single laser powder bed fusion (LPBF) printed cobalt-chromium alloy (Co28Cr6Mo) part. The external surface of the implant that is intended to be in contact with the phalanx is mirror polished. Component: Not a component
Reason / hazard
Implants were found in the field with orange sterilzation indicators which means implants were not exposed to sterilization.
What to do (remedy)

How to claim your remedy

Classification
Class II
Status
Ongoing
Distribution / sold at
U.S. Nationwide distribution in the states of FL, IL, NC, PA, SC, and TX.
Recall initiated
2026-09-04
Code info
Lot Code: Model: 1103-161620 UDI: 00840097505244 Lot: A02979
Product quantity
16
Voluntary mandated
Voluntary: Firm initiated
City
Durham
State
NC
Country
United States
Event id
99940
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.