⚠ RecallRadar📷 ScanGuidesRecentOfficial U.S. recall notices, restated plainly

Greiner Bio-One North America, Inc.: Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood

Medical DevicesSource: FDA (device) · Date: 2026-10-28

Recall status
Ongoing (still active)
Severity
Class II: could cause temporary or reversible health problems What recall classes mean
Where sold
U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA.

Common questions

Is Greiner Bio-One North America LH Lithium Heparin Separator recalled?
Yes. FDA (device) recalled Greiner Bio-One North America LH Lithium Heparin Separator on 2026-10-28. Recall status: Ongoing (still active).
What is the hazard?
Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.
Where was it sold?
U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA.

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Is Greiner Bio-One North America recalled?

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

More recalls

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Greiner Bio-One North America, Inc.
Product
Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml)
Reason / hazard
Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.
What to do (remedy)

How to claim your remedy

Classification
Class II
Status
Ongoing
Distribution / sold at
U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA.
Recall initiated
2026-07-29
Code info
Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12
Product quantity
860450
Voluntary mandated
Voluntary: Firm initiated
City
Monroe
State
NC
Country
United States
Event id
99518
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.