Medline Industries, LP: Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135 LAP CHOLE DYNJ911660 LAP NISSEN
- Recall status
- Ongoing (still active)
- Severity
- Class II: could cause temporary or reversible health problems What recall classes mean
- Where sold
- US Nationwide distribution in the states of AZ, CA, IN, MD, MO, NJ, OH, and WI.
Common questions
- Is Medline Industries Convenience Kits containing CONMED AirSeal Access Ports recalled?
- Yes. FDA (device) recalled Medline Industries Convenience Kits containing CONMED AirSeal Access Ports on 2026-10-21. Recall status: Ongoing (still active).
- What is the hazard?
- Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
- Where was it sold?
- US Nationwide distribution in the states of AZ, CA, IN, MD, MO, NJ, OH, and WI.
Read the official notice at FDA (device) →
Is Medline Industries recalled?
More recalls
- Medline Industries, LP: MEDLINE medical convenience kits labeled as: 1) HIP
- Medline Industries, LP: MEDLINE medical convenience kits labeled as: C-SECTION
- Medline Industries, LP: MEDLINE medical convenience kits labeled as: TUBAL
- Medline Industries, LP: MEDLINE medical convenience kits labeled as: (list
- Medline Industries, LP: MEDLINE medical convenience kits labeled as: 1)
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medline Industries, LP
- Product
- Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135 LAP CHOLE DYNJ911660 LAP NISSEN DYNJ911560 UROLOGY LAPAROSCOPIC - 51638 DYNJ908720A
- Reason / hazard
- Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
- What to do (remedy)
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the states of AZ, CA, IN, MD, MO, NJ, OH, and WI.
- Recall initiated
- 2026-08-31
- Code info
- DYNJ911135 UDI-DI 10198459264382 Lots 25EDA279 25FDB612 25LDB658 DYNJ911660 UDI-DI 10198459487071 Lots 25IMF444 25JMA458 26BMG519 26DMA914 26EMA779 26FME775 26GMA849 26GMF826 26GMM101 26GMM127 DYNJ911560 UDI-DI 10198459400261 Lots 25HMG749 25JMC886 25KMA383 25KMA392 DYNJ908720A UDI-DI 10195327423155 Lots 24DBN292 24GBJ114 24IBK504 24JBC471 25KBP733 26ABG732
- Product quantity
- 3,551 kits
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Northfield
- State
- IL
- Country
- United States
- Event id
- 99888