⚠ RecallRadar📷 ScanGuidesOfficial U.S. recall notices, restated plainly

American Contract Systems Inc.: Medical convenience kits Item Number/Description ANCV78BC HEART PACK - 205947 HGCV05K CV SET UP PACK MHPP97AR PACEMAKER PACK- 206010

Medical DevicesSource: FDA (device) · Date: 2025-10-22

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for American Contract Systems

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
American Contract Systems Inc.
Product
Medical convenience kits Item Number/Description ANCV78BC HEART PACK - 205947 HGCV05K CV SET UP PACK MHPP97AR PACEMAKER PACK- 206010 UDBH37AW BASIC HEART PACK - 206044 UHPP82AH PERIPHERAL PACK - 209533 UICD62AN CARDIOVASCUALR SUPPLY (PS 907285) UICT16F CARDIO THORACIC DRAPE PACK (PS 907284) UIVA75I VASCULAR PACK
Reason / hazard
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Recall initiated
2025-09-04
Code info
ANCV78BB UDI-DI 191072214362 Lot 8133811 Bag serial number 68273380; ANCV78BC UDI-DI 191072235169 Lot 8499811 Bag serial numbers N/A; HGCV05K UDI-DI 191072212474 Lot 8021411 Bag serial number 68154464; MHPP97AR UDI-DI 191072237446 Lot 8527511 Bag serial number N/A; UDBH37AW UDI-DI 191072235299 Lot 8194311 Bag serial numbers 68198334 68198335; UHPP82AH UDI-DI 191072216434 Lot 8112811 Bag serial number N/A; UICD62AN UDI-DI 191072200495 Lot 7999011 Bag serial number N/A; UICT16F UDI-DI 191072150431 Lot 8000411 Bag serial number N/A Lot 8331311 Bag serial number 17627819 Lot 8429011 Bag serial number 17749736; UIVA75I UDI-DI 191072229052 Lot 8056011 Bag serial number N/A; UIVA75J UDI-DI 191072235602 Lot 8422011 Bag serial number 68667536 * If Bag serial number is "n/a", then all of that lot is affected
Product quantity
32,433 total
Voluntary mandated
Voluntary: Firm initiated
City
Grimes
State
IA
Country
United States
Event id
97659
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.