Steris Corporation: RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115
- Recall status
- Ongoing (still active)
- Severity
- Class II: could cause temporary or reversible health problems What recall classes mean
- Where sold
- Worldwide - US Nationwide distribution.
Common questions
- Is Steris RAPICIDE PA Part A High-Level Disinfectant recalled?
- Yes. FDA (device) recalled Steris RAPICIDE PA Part A High-Level Disinfectant on 2026-10-21. Recall status: Ongoing (still active).
- What is the hazard?
- If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is...
- Where was it sold?
- Worldwide - US Nationwide distribution.
Read the official notice at FDA (device) →
More recalls
- Steris Corporation: RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog
- Steris Corporation: RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog
- Steris Corporation: HexaVue Focus Integration System, a single, all-in-one
- Steris Corporation: VERIFY Biological Steam Test Pack Model/Catalog Number
- Steris Corporation: VERIFY Biological Steam Test Pack Model/Catalog Number
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Steris Corporation
- Product
- RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115
- Reason / hazard
- If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
- What to do (remedy)
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution.
- Recall initiated
- 2026-09-21
- Code info
- UDI: 00724995245658 Lot numbers with expiration dates: 687046 10/5/2027 688017 10/19/2027
- Product quantity
- 374 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Mentor
- State
- OH
- Country
- United States
- Event id
- 99804
Free daily digest + "My products" watchlist