Philips Ultrasound, LLC: Philips Lumify Diagnostic Ultrasound System
- Recall status
- Ongoing (still active)
- Severity
- Class I: most serious, could cause serious health problems or death What recall classes mean
- Where sold
- Worldwide - US Nationwide distribution including in the states of CA, MA, IL, NY, OR, GA, TX, WA, FL, PA, WI, LA, NJ, KY, VA, MS, AZ, SC, CT, MN, UT, OH, AL, MI, NH, MO, TN, IN, DC, NE, NM, MT, NV...
Common questions
- Is Philips Ultrasound Lumify Diagnostic System recalled?
- Yes. FDA (device) recalled Philips Ultrasound Lumify Diagnostic System on 2026-10-28. Recall status: Ongoing (still active).
- What is the hazard?
- Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable...
- Where was it sold?
- Worldwide - US Nationwide distribution including in the states of CA, MA, IL, NY, OR, GA, TX, WA, FL, PA, WI, LA, NJ, KY, VA, MS, AZ, SC, CT, MN, UT, OH, AL, MI, NH, MO, TN, IN, DC, NE, NM, MT, NV...
Read the official notice at FDA (device) →
Is Philips Ultrasound recalled?
More recalls
- Philips Ultrasound, LLC: Philips EPIQ Elite Diagnostic Ultrasound System Model #
- Philips Ultrasound, Inc: Philips X7-2 Ultrasound Transducer
- Philips Ultrasound, Inc: Philips X3-1 Ultrasound Transducer
- Philips Ultrasound, Inc: Philips S5-2 Ultrasound Transducer
- Philips Ultrasound, Inc: Philips S4-1 Ultrasound Transducer
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Philips Ultrasound, LLC
- Product
- Philips Lumify Diagnostic Ultrasound System
- Reason / hazard
- Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable connection/may disconnect disrupting function. May cause diagnosis/treatment delay.
- What to do (remedy)
- Classification
- Class I
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution including in the states of CA, MA, IL, NY, OR, GA, TX, WA, FL, PA, WI, LA, NJ, KY, VA, MS, AZ, SC, CT, MN, UT, OH, AL, MI, NH, MO, TN, IN, DC, NE, NM, MT, NV, NC, MD, IA, SD, VT, KS, AK, CO, ID, OK, WV, AR, RI, ME, WY, HI, DE, ND and the countries of CA, NO, HR, DE, KR, GB, ES, FI, PL, AU, TH, SG, NL, RO, CH, FR, MM, MX, BE, DK, MY, SE, LB, SI, BR, CL, GR, PH, CN, ID, LT, HK, MQ, JP, IT, IE, IL, AT, TW, AE, SA, HU, RU, VN, NZ, TR, SK, ZA, KE, AR, NP, PK, UY, DO, MN, LK, GT, BD, IS, IN, NI, LY, CY, CZ, SV, PA, BG, PE, EC, CO, VE, UZ, YE, KW, DZ, LU, EG, EE, UA, LV, KZ, TZ, MD, JO, GY, QA, MU, AM, OM, PT, TN, NG, BO, PF, ZM, GF, YT, ME, GD, RS, BW, GP, RE, MT, CR, MA, CI, NC, BA, SR, CG, TL, GH, AL, MV, XK, BN, LS, MK, CM, TT, ZW, BH, PS, HN, MC, WS, JE, PM, AZ.
- Recall initiated
- 2026-09-24
- Code info
- UDI-DI/Software Version: Android: 00884838122116/1.X, 3.X, 4.X, 5.X. iOS: 00884838124141/2.X and 5.X
- Product quantity
- 13,206
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Bothell
- State
- WA
- Country
- United States
- Event id
- 99878
Free daily digest + "My products" watchlist