Medtronic Sofamor Danek USA Inc: Grafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive to yield a demineralized bone mat
- Recall status
- Ongoing (still active)
- Severity
- Class II: could cause temporary or reversible health problems What recall classes mean
- Where sold
- US Nationwide distribution in the states of OK, VA, CO, AR, FL, TX, MD, DE, IL, MI, KS, AL, GA.
Common questions
- Is Medtronic Sofamor Danek Grafton" DBM and Grafton Plus" DBM contain demineralized recalled?
- Yes. FDA (device) recalled Medtronic Sofamor Danek Grafton" DBM and Grafton Plus" DBM contain demineralized on 2026-10-21. Recall status: Ongoing (still active).
- What is the hazard?
- Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product sterility testing Medtronic is retrieving any unused product manufactured...
- Where was it sold?
- US Nationwide distribution in the states of OK, VA, CO, AR, FL, TX, MD, DE, IL, MI, KS, AL, GA.
Read the official notice at FDA (device) →
Is Medtronic Sofamor Danek recalled?
More recalls
- Medtronic Sofamor Danek USA Inc: Grafton" DBM and Grafton Plus" DBM contain
- Medtronic Sofamor Danek USA Inc: Catalyft PL & PL40 Expandable Interbody System
- Medline: Beds, Rails, Mattresses, and accessories
- ARJOHUNTLEIGH POLSKA Sp. z.o.o.: Carevo, Height-adjustable shower trolley, Model
- Stryker Leibinger GmbH & Co. KG: Precision Pointer Electromagnetic
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medtronic Sofamor Danek USA Inc
- Product
- Grafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive to yield a demineralized bone matrix (DBM) allograft product having a particular physical form and/or handling property. Grafton DBM Gel and Grafton Plus DBM Paste are produced from a powder form of DBM, while Grafton DBM Flex, Putty, Matrix, Crunch" and Orthoblend are produced from a fiber form of DBM Crunch" also contains demineralized bone chips/cubes, while Grafton DBM Orthoblend contains non-demineralized cancellous bone chips in addition to DBM. Grafton DBM and Grafton Plus DBM are malleable and can be molded or cut into various sizes and shapes according to the intended implant site.
- Reason / hazard
- Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product sterility testing Medtronic is retrieving any unused product manufactured from the donor tissue. Donor bone tissue tested positive for Clostridium butyricum and while tissue did undergo gamma radiation, if implanted, there is a possibility of clostridium infection.
- What to do (remedy)
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the states of OK, VA, CO, AR, FL, TX, MD, DE, IL, MI, KS, AL, GA.
- Recall initiated
- 2026-08-12
- Code info
- UDI-DI: 00643169122932. Serial Numbers: A62881-014 through A62881-033, specifically: A62881-014, A62881-015, A62881-016, A62881-017, A62881-018, A62881-019, A62881-020, A62881-021, A62881-022, A62881-023, A62881-024, A62881-025, A62881-026, A62881-027, A62881-028, A62881-029, A62881-030, A62881-031, A62881-032, A62881-033
- Product quantity
- 20
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Memphis
- State
- TN
- Country
- United States
- Event id
- 99701