POLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC PRODUCTS, LLC: SEDIPLAST; Model/Catalog Numbers: S-100 and S-1000;
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- POLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC PRODUCTS, LLC
- Product
- SEDIPLAST; Model/Catalog Numbers: S-100 and S-1000;
- Reason / hazard
- During use of an affected tube, the operator may encounter increased resistance when inserting the pipette. The resulting pressure buildup may cause the tube cap to detach or be forcibly ejected, potentially resulting in the release of blood and exposure of the operator or surrounding work area to a biological hazard (blood).
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution in the states and territory of AK, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV and the country of United Kingdom.
- Recall initiated
- 2026-08-25
- Code info
- Model/Catalog No. S-100 and S-1000; UDI-DI: N/A; Lot Number E1150A; Expiration Date 06/30/27
- Product quantity
- 448400
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Cortlandt Manor
- State
- NY
- Country
- United States
- Event id
- 99810