⚠ RecallRadar📷 ScanGuidesOfficial U.S. recall notices, restated plainly

B BRAUN MEDICAL INC: Brand Name: Hemodialysis Bloodlines Product Name: STREAMLINE BLOODLINE SET FOR DIALOG Model/Catalog Number: SL-2010M2096 Software Version

Medical DevicesSource: FDA (device) · Date: 2025-10-29

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for B Braun Medical

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
B BRAUN MEDICAL INC
Product
Brand Name: Hemodialysis Bloodlines Product Name: STREAMLINE BLOODLINE SET FOR DIALOG Model/Catalog Number: SL-2010M2096 Software Version: N/A Product Description: STREAMLINE BLOODLINE SET FOR DIALOG Component: N/A
Reason / hazard
The potential for micro-air bubbles observed in the bloodline and air-in-line alarms due to damaged arterial and venous patient connectors.
Classification
Class I
Status
Ongoing
Distribution / sold at
Domestic: US Nationwide Distribution.
Recall initiated
2025-09-30
Code info
Model No. SL-2010M2096; UDI-DI 04046964367786; Lot Numbers: A2500119, A2500121, A2500128, A2500129, A2500130, A2500132, A2500133, A2500134, A2500149, A2500150, A2500151, A2500180, A2500181, A2500182, A2500185, A2500198, A2500199, A2500200, A2500203, A2500204, A2500205, A2500206, A2500207, A2500208, A2500209, A2500210, A2500211, A2500212, A2500213, A2500214, A2500215, A2500216, A2500217, A2500218, A2500219, A2500220, A2500246, A2500247, A2500248, A2500249, A2500250, A2500251, A2500257, A2500258, A2500259, A2500260, A2500261, A2500262, A2500267, A2500268, A2500269, A2500273, A2500274, A2500275, A2500276, A2500277, A2500278, A2500295, A2500298, A2500299, A2500300; ***Addded 3/24/2026***Additional Lots Identified A2500318, A2500319, A2500320, A2500321, A2500322, A2500323, A2500324, A2500325, A2500326, A2500327, A2500328, A2500329, A2500330, A2500356, A2500357;
Product quantity
1298454 units (1589674 - 3/5/26 Expansion)
Voluntary mandated
Voluntary: Firm initiated
City
Allentown
State
PA
Country
United States
Event id
97609
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.